Recall Z-2467-2018— GE Medical Systems Ultrasound & Primary Care Diagnostics, LL
Class II · Terminated
Class II FDA medical device recall by GE Medical Systems Ultrasound & Primary Care Diagnostics, LL, initiated 2018-04-02, terminated 2024-04-30. FDA's recall record names no specific 510(k) clearance for this event. Product: T2100 Micro flex Drive Treadmill, powered Product Usage: Powered exercise equipment consist of powered devices intended for medical purposes, such as to redevelop muscles or restore motion to joints or for use as an adjunct treatment for obesity. Specifically the T210 Treadmill is intended for use in Exercise Testing, facilitating accurate blood pressure measurements and exercise testing within speed range of 0-13.5 miles per hour.. Reason: A performance issue with customer owned spare parts, T2100 Microflex drive (2026182-002 or 2026182-004), was not addressed with a previous safety correction. If these parts were installed from customer owned stock on the T2100 Treadmill, uncontrolled walking belt motion during a stress exercise test could occur..
- Recalling firm
- GE Medical Systems Ultrasound & Primary Care Diagnostics, LL
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-04-02
- Terminated
- 2024-04-30
- Firm location
- Milwaukee, WI, United States
Product description
T2100 Micro flex Drive Treadmill, powered Product Usage: Powered exercise equipment consist of powered devices intended for medical purposes, such as to redevelop muscles or restore motion to joints or for use as an adjunct treatment for obesity. Specifically the T210 Treadmill is intended for use in Exercise Testing, facilitating accurate blood pressure measurements and exercise testing within speed range of 0-13.5 miles per hour.
Reason for recall
A performance issue with customer owned spare parts, T2100 Microflex drive (2026182-002 or 2026182-004), was not addressed with a previous safety correction. If these parts were installed from customer owned stock on the T2100 Treadmill, uncontrolled walking belt motion during a stress exercise test could occur.
Data sourced from openFDA. This site is unofficial and independent of the FDA.