Recall Z-2462-2017— Biomerieux Inc

Class II · Terminated

Class II FDA medical device recall by Biomerieux Inc, initiated 2017-04-20, terminated 2020-07-22. FDA's recall record names no specific 510(k) clearance for this event. Product: VITEK¿ 2 Gram Positive Susceptibility card (AST-GP74), REF 414971, 20 cards per carton.. Reason: The integrity of the product container may be compromised thereby exposing product to moisture which could lead to antibiotic degradation (loss of potency)..

Recalling firm
Biomerieux Inc
Classification
Class II
Status
Terminated
Initiated
2017-04-20
Terminated
2020-07-22
Firm location
Hazelwood, MO, United States

Product description

VITEK¿ 2 Gram Positive Susceptibility card (AST-GP74), REF 414971, 20 cards per carton.

Reason for recall

The integrity of the product container may be compromised thereby exposing product to moisture which could lead to antibiotic degradation (loss of potency).