Recall Z-2459-2024— GE Medical Systems, LLC

Class II · Ongoing

Class II FDA medical device recall by GE Medical Systems, LLC, initiated 2024-06-14. It names 3 FDA 510(k) clearances: K202966, K163331, K160618. Product: SIGNA Architect, with affected software versions: DV26.0 to DV26.3, DV27.0 to DV27.3, DV28.0 to DV28.5, DV29.0, DV29.1, DV29.2, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System. Reason: GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant..

Recalling firm
GE Medical Systems, LLC
Classification
Class II
Status
Ongoing
Initiated
2024-06-14
Firm location
Waukesha, WI, United States

510(k) clearances named in this recall

3 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

SIGNA Architect, with affected software versions: DV26.0 to DV26.3, DV27.0 to DV27.3, DV28.0 to DV28.5, DV29.0, DV29.1, DV29.2, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System

Reason for recall

GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.

Data sourced from openFDA. This site is unofficial and independent of the FDA.