Recall Z-2457-2019— Allergan PLC
Class I · Ongoing
Class I FDA medical device recall by Allergan PLC, initiated 2019-07-24. It names 2 FDA 510(k) clearances: K102806, K862203. Product: Natrelle 133 Tissue Expanders w/Suture Tabs BIOCELL Full Height Variable Projection with MAGNA-FINDEr Xact & 21G Needle Infusion Set STYLE 133FV-T: 133FV-11-T, 133FV-12-T, 133FV-13-T, 133FV-14-T, 133FV-15-T, 133FV-16-T; STYLE 133MV-T: 133MV-11-T, 133MV-12-T, 133MV-13-T, 133MV-14-T, 133MV-15-T, 133MV-16-T; STYLE 133LV-T: 133LV-11-T, 133LV-12-T, 133LV-13-T, 133LV-14-T, 133LV-15-T, 133LV-16-T; STYLE 133SV-T: 133SV-11-T, 133SV-12-T, 133SV-13-T, 133SV-14-T, 133SV-15-T, 133SV-16-T STYLE 133FX-T: 133FX-11-T, 133FX-12-T, 133FX-13-T, 133FX-14-T, 133FX-15-T, 133FX-16-T; STYLE 133MX-T: 133MX-11-T, 133MX-12-T, 133MX-13-T, 133MX-14-T, 133MX-15-T, 133MX-16-T; STYLE 133SX-T: 133SX-11-T, 133SX-12-T, 133SX-13-T, 133SX-14-T, 133SX-15-T, 133SX-16-T; Product Usage: intended for temporary subcutaneous implantation and require periodic, incremental inflation with sterile saline for injection until the desired amount of tissue is developed.. Reason: The action was initiated following notification by the U.S. FDA of their recently updated, global safety information concerning the higher incidence of anaplastic large cell lymphoma (BIA-ALCL) in patients who have textured breast implants..
- Recalling firm
- Allergan PLC
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2019-07-24
- Firm location
- Irvine, CA, United States
510(k) clearances named in this recall
2 on FDA's record- K102806 — NATRELLE 133 TISSUE EXPANDER WITH SUTURE TABS
- K862203 — MCGHAN MAGNA-SITE(TM) TISSUE EXPANDER
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Natrelle 133 Tissue Expanders w/Suture Tabs BIOCELL Full Height Variable Projection with MAGNA-FINDEr Xact & 21G Needle Infusion Set STYLE 133FV-T: 133FV-11-T, 133FV-12-T, 133FV-13-T, 133FV-14-T, 133FV-15-T, 133FV-16-T; STYLE 133MV-T: 133MV-11-T, 133MV-12-T, 133MV-13-T, 133MV-14-T, 133MV-15-T, 133MV-16-T; STYLE 133LV-T: 133LV-11-T, 133LV-12-T, 133LV-13-T, 133LV-14-T, 133LV-15-T, 133LV-16-T; STYLE 133SV-T: 133SV-11-T, 133SV-12-T, 133SV-13-T, 133SV-14-T, 133SV-15-T, 133SV-16-T STYLE 133FX-T: 133FX-11-T, 133FX-12-T, 133FX-13-T, 133FX-14-T, 133FX-15-T, 133FX-16-T; STYLE 133MX-T: 133MX-11-T, 133MX-12-T, 133MX-13-T, 133MX-14-T, 133MX-15-T, 133MX-16-T; STYLE 133SX-T: 133SX-11-T, 133SX-12-T, 133SX-13-T, 133SX-14-T, 133SX-15-T, 133SX-16-T; Product Usage: intended for temporary subcutaneous implantation and require periodic, incremental inflation with sterile saline for injection until the desired amount of tissue is developed.
Reason for recall
The action was initiated following notification by the U.S. FDA of their recently updated, global safety information concerning the higher incidence of anaplastic large cell lymphoma (BIA-ALCL) in patients who have textured breast implants.
Data sourced from openFDA. This site is unofficial and independent of the FDA.