Recall Z-2456-2024— BioFire Diagnostics, LLC
Class II · Ongoing
Class II FDA medical device recall by BioFire Diagnostics, LLC, initiated 2024-06-19. It names one FDA 510(k) clearance: K193519. Product: BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, for FILMARRAY Systems, when used in conjunction with Culture Media Bottles: BACT/ALERT FA PLUS, REF: 410851, and BACT/ALERT PF PLUS, REF: 410853. Reason: If blood culture identification panel is used in conjunction with specific lots of culture media bottles, then false positive Serratia marcescens results may occur, due to an increased level of non-viable organism from serratia marcescens targets in culture media bottles, false positive result may lead to an inappropriate change in patient therapy..
- Recalling firm
- BioFire Diagnostics, LLC
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2024-06-19
- Firm location
- Salt Lake City, UT, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, for FILMARRAY Systems, when used in conjunction with Culture Media Bottles: BACT/ALERT FA PLUS, REF: 410851, and BACT/ALERT PF PLUS, REF: 410853
Reason for recall
If blood culture identification panel is used in conjunction with specific lots of culture media bottles, then false positive Serratia marcescens results may occur, due to an increased level of non-viable organism from serratia marcescens targets in culture media bottles, false positive result may lead to an inappropriate change in patient therapy.