Recall Z-2449-2012— Terumo Cardiovascular Systems Corporation

Class II · Terminated

Class II FDA medical device recall by Terumo Cardiovascular Systems Corporation, initiated 2012-09-20, terminated 2014-01-13. It names one FDA 510(k) clearance: K022947. Product: Terumo Advanced Perfusion System 1 Flow Module indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment.. Reason: Terumo Cardiovascular Systems (CVS) received one report of a flow module for the Terumo Advanced Perfusion System 1 not displaying blood flow information on the Central Control Monitor or the Centrifugal Control Unit. The tricolor LED on the flow module was not illuminated. Terumo CVS' investigation found that the malfunction was caused by an improper component on a circuit board used in a populat.

Recalling firm
Terumo Cardiovascular Systems Corporation
Classification
Class II
Status
Terminated
Initiated
2012-09-20
Terminated
2014-01-13
Firm location
Ann Arbor, MI, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Terumo Advanced Perfusion System 1 Flow Module indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment.

Reason for recall

Terumo Cardiovascular Systems (CVS) received one report of a flow module for the Terumo Advanced Perfusion System 1 not displaying blood flow information on the Central Control Monitor or the Centrifugal Control Unit. The tricolor LED on the flow module was not illuminated. Terumo CVS' investigation found that the malfunction was caused by an improper component on a circuit board used in a populat

Data sourced from openFDA. This site is unofficial and independent of the FDA.