Recall Z-2445-2024— LINK BIO CORP

Class II · Ongoing

Class II FDA medical device recall by LINK BIO CORP, initiated 2024-06-27. FDA's recall record names no specific 510(k) clearance for this event. Product: LINK SymphoKnee Patella Sizing Template, Item Number 881-509/00. Reason: The LinkSymphoKnee Patella Sizing Template was incorrectly labeled. The markings on two arms of the instrument were mixed up because the triangular Patella Sizing Template was inserted upside down into the laser marking process..

Recalling firm
LINK BIO CORP
Classification
Class II
Status
Ongoing
Initiated
2024-06-27
Firm location
Dover, NJ, United States

Product description

LINK SymphoKnee Patella Sizing Template, Item Number 881-509/00

Reason for recall

The LinkSymphoKnee Patella Sizing Template was incorrectly labeled. The markings on two arms of the instrument were mixed up because the triangular Patella Sizing Template was inserted upside down into the laser marking process.