Recall Z-2441-2018— Zeiss, Carl Inc
Class II · Terminated
Class II FDA medical device recall by Zeiss, Carl Inc, initiated 2018-05-29, terminated 2024-07-26. FDA's recall record names no specific 510(k) clearance for this event. Product: Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope. Reason: Under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in exposure of the users to harmful UV radiation..
- Recalling firm
- Zeiss, Carl Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-05-29
- Terminated
- 2024-07-26
- Firm location
- Thornwood, NY, United States
Product description
Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope
Reason for recall
Under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in exposure of the users to harmful UV radiation.