Recall Z-2437-2018— Siemens Healthcare Diagnostics, Inc.

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2018-03-21, terminated 2020-06-12. FDA's recall record names no specific 510(k) clearance for this event. Product: Dimension Vista¿ MULTI 1 SDIL, KD693, SMN# 10469971. Reason: The diluent may have an incomplete slit on the septum in the cap of the vial. A diluent cap with a defective septum has the potential to cause erroneously elevated results if a sample is autodiluted onboard the system, due to a reduced volume of diluent being pipetted from the vial..

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Terminated
Initiated
2018-03-21
Terminated
2020-06-12
Firm location
Newark, DE, United States

Product description

Dimension Vista¿ MULTI 1 SDIL, KD693, SMN# 10469971

Reason for recall

The diluent may have an incomplete slit on the septum in the cap of the vial. A diluent cap with a defective septum has the potential to cause erroneously elevated results if a sample is autodiluted onboard the system, due to a reduced volume of diluent being pipetted from the vial.