Recall Z-2427-2024— Medtronic Neuromodulation

Class II · Ongoing

Class II FDA medical device recall by Medtronic Neuromodulation, initiated 2024-05-15. FDA's recall record names no specific 510(k) clearance for this event. Product: A610 Clinician Programmer (CP) Software Application (app) with the following versions: Version 2.0.x (2.0.4584, 2.0.4594, 2.0.4605, 2.0.4630, and 2.0.4648), Version 3.0.x (3.0.1057, 3.0.1062, 3.0.1081, and 3.0.1098), Version 4.0.x (4.0.1052). Includes the below product descriptions: 1. ADAPTOR 64001 1X4 MVMT POCKET ADAPTOR, 2. ADAPTOR 64002 2X4 MVMT POCKET ADAPTOR, 3. INS 37602 Activa SC Neurostimulator, 4. INS 37603 Activa SCx Neurostimulator, 5. INS B35200 DBS PERCEPT PC, 6. INS B35300 PERCEPT RC V1. Reason: Patients implanted with a pocket adaptor (Model 64001 and/or 64002) are limited to HEAD ONLY MRI eligibility..

Recalling firm
Medtronic Neuromodulation
Classification
Class II
Status
Ongoing
Initiated
2024-05-15
Firm location
Minneapolis, MN, United States

Product description

A610 Clinician Programmer (CP) Software Application (app) with the following versions: Version 2.0.x (2.0.4584, 2.0.4594, 2.0.4605, 2.0.4630, and 2.0.4648), Version 3.0.x (3.0.1057, 3.0.1062, 3.0.1081, and 3.0.1098), Version 4.0.x (4.0.1052). Includes the below product descriptions: 1. ADAPTOR 64001 1X4 MVMT POCKET ADAPTOR, 2. ADAPTOR 64002 2X4 MVMT POCKET ADAPTOR, 3. INS 37602 Activa SC Neurostimulator, 4. INS 37603 Activa SCx Neurostimulator, 5. INS B35200 DBS PERCEPT PC, 6. INS B35300 PERCEPT RC V1

Reason for recall

Patients implanted with a pocket adaptor (Model 64001 and/or 64002) are limited to HEAD ONLY MRI eligibility.

Data sourced from openFDA. This site is unofficial and independent of the FDA.