Recall Z-2423-2018— Anika Therapeutics, Inc.

Class II · Terminated

Class II FDA medical device recall by Anika Therapeutics, Inc., initiated 2018-05-10, terminated 2019-05-10. It names one FDA 510(k) clearance: K073251. Product: Hyalomatrix, Product nos. 651105, 651103, 651104, and 651301 Product Usage: Hyalomatrix is a bi-layered, sterile, flexible, and comfortable wound dressing that acts as an advanced wound care device. It is comprised of a non-woven pad entirely composed of HY AFF 11, a benzyl ester of hyaluronic acid, and a semipermeable silicone membrane, which controls water vapor loss, provides a flexible covering for the wound surface, and adds increased tear strength to the device. The HYAFFl 1 wound contact layer biodegradable matrix acts as a scaffold for cellular invasion and capillary growth. Hyalomatrix is indicated for the management of wounds including: partial and full-thickness wounds; second-degree bums; pressure ulcers; venous ulcers; diabetic ulcers; chronic vascular ulcers; tunneled/undetermined wounds; surgical wounds (donor sites/grafts, post-Mohs surgery, post-laser surgery, podiatric, wound dehiscence); trauma wounds (abrasions, lacerations, skin tears); and draining wounds. The device is intended for one-time use.. Reason: Hyalomatrix products are susceptible to degradation as a result of the decomposition of PVC in the packaging material during sterilization, which resulted in degradation of the film. Over the long term it is possible that product integrity may be compromised..

Recalling firm
Anika Therapeutics, Inc.
Classification
Class II
Status
Terminated
Initiated
2018-05-10
Terminated
2019-05-10
Firm location
Bedford, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Hyalomatrix, Product nos. 651105, 651103, 651104, and 651301 Product Usage: Hyalomatrix is a bi-layered, sterile, flexible, and comfortable wound dressing that acts as an advanced wound care device. It is comprised of a non-woven pad entirely composed of HY AFF 11, a benzyl ester of hyaluronic acid, and a semipermeable silicone membrane, which controls water vapor loss, provides a flexible covering for the wound surface, and adds increased tear strength to the device. The HYAFFl 1 wound contact layer biodegradable matrix acts as a scaffold for cellular invasion and capillary growth. Hyalomatrix is indicated for the management of wounds including: partial and full-thickness wounds; second-degree bums; pressure ulcers; venous ulcers; diabetic ulcers; chronic vascular ulcers; tunneled/undetermined wounds; surgical wounds (donor sites/grafts, post-Mohs surgery, post-laser surgery, podiatric, wound dehiscence); trauma wounds (abrasions, lacerations, skin tears); and draining wounds. The device is intended for one-time use.

Reason for recall

Hyalomatrix products are susceptible to degradation as a result of the decomposition of PVC in the packaging material during sterilization, which resulted in degradation of the film. Over the long term it is possible that product integrity may be compromised.