Recall Z-2421-2018— Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class II · Terminated
Class II FDA medical device recall by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), initiated 2018-06-21, terminated 2021-06-16. FDA's recall record names no specific 510(k) clearance for this event. Product: Medtronic EnTrust Escudo D144DRG (OUS distribution) Product Usage: The implantable cardioverter defibrillator is intended to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias.. Reason: Medtronic has become aware of a potential for loss of high voltage and anti-tachycardia pacing therapy in EnTrust and Escudo implantable cardioverter defibrillators (ICDs) as they near elective replacement indicator (ERI) voltage. Under certain circumstances, the device may display an immediate End of Life (EOL) Observation with no prior ERI alert. Though no ERI alert is triggered, there may not be enough remaining battery capacity to charge the high voltage circuits, resulting in an excessive charge time EOL Observation, leading to a loss of high voltage and anti-tachycardia pacing therapy..
- Recalling firm
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-06-21
- Terminated
- 2021-06-16
- Firm location
- Mounds View, MN, United States
Product description
Medtronic EnTrust Escudo D144DRG (OUS distribution) Product Usage: The implantable cardioverter defibrillator is intended to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias.
Reason for recall
Medtronic has become aware of a potential for loss of high voltage and anti-tachycardia pacing therapy in EnTrust and Escudo implantable cardioverter defibrillators (ICDs) as they near elective replacement indicator (ERI) voltage. Under certain circumstances, the device may display an immediate End of Life (EOL) Observation with no prior ERI alert. Though no ERI alert is triggered, there may not be enough remaining battery capacity to charge the high voltage circuits, resulting in an excessive charge time EOL Observation, leading to a loss of high voltage and anti-tachycardia pacing therapy.
Data sourced from openFDA. This site is unofficial and independent of the FDA.