Recall Z-2418-2020— Hocoma AG
Class II · Ongoing
Class II FDA medical device recall by Hocoma AG, initiated 2020-06-08. FDA's recall record names no specific 510(k) clearance for this event. Product: Lokomat Pro, Catalog Numbers 30003 LokomatPro L6.2 120V AD ; 30004 LokomatPro L6.2 120V Kombi; 30005 LokomatPro L6.2 120V PE; 30683 LokomatPro L6.2 110V FD PE; 30684 LokomatPro L6.2 110V FD Kombi; 32001 LokomatPro L6.2 110V FD AD Product Usage: The intended use of the Lokomat is to support treadmill training to treat patients with walking disabilities caused by neurological, muscular or bone-related disorders. isokinetic testing and evaluation system. Reason: The motor controller may fail, which can potentially lead to an error in functionality of the Body Weights Support Rope, which means that the rope can go up or down in an uncontrolled way..
- Recalling firm
- Hocoma AG
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2020-06-08
- Firm location
- Volketswil, N/A, Switzerland
Product description
Lokomat Pro, Catalog Numbers 30003 LokomatPro L6.2 120V AD ; 30004 LokomatPro L6.2 120V Kombi; 30005 LokomatPro L6.2 120V PE; 30683 LokomatPro L6.2 110V FD PE; 30684 LokomatPro L6.2 110V FD Kombi; 32001 LokomatPro L6.2 110V FD AD Product Usage: The intended use of the Lokomat is to support treadmill training to treat patients with walking disabilities caused by neurological, muscular or bone-related disorders. isokinetic testing and evaluation system
Reason for recall
The motor controller may fail, which can potentially lead to an error in functionality of the Body Weights Support Rope, which means that the rope can go up or down in an uncontrolled way.
Data sourced from openFDA. This site is unofficial and independent of the FDA.