Class II · Ongoing
Class II FDA medical device recall by EVIDENT SCIENTIFIC INC, initiated 2024-07-22. FDA's recall record names no specific 510(k) clearance for this event. Product: SlideView VS200 slide viewer,. Reason: A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated..
- Recalling firm
- EVIDENT SCIENTIFIC INC
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2024-07-22
- Firm location
- Waltham, MA, United States
Product description
SlideView VS200 slide viewer,
Reason for recall
A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated.