Recall Z-2415-2025— Baxter Healthcare Corporation
Class II · Ongoing
Class II FDA medical device recall by Baxter Healthcare Corporation, initiated 2025-07-14. It names one FDA 510(k) clearance: K230041. Product: Spectrum IQ Infusion Pump, Product Code 3570009. Reason: Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy..
- Recalling firm
- Baxter Healthcare Corporation
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-07-14
- Firm location
- Deerfield, IL, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Spectrum IQ Infusion Pump, Product Code 3570009
Reason for recall
Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy.
Data sourced from openFDA. This site is unofficial and independent of the FDA.