Recall Z-2404-2021— Siemens Medical Solutions USA, Inc

Class II · Terminated

Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2021-06-30, terminated 2023-02-10. It names one FDA 510(k) clearance: K181407. Product: Artis Models with SW version VD12 listed below: Artis zee floor 10094135 Artis zee ceiling 10094137 Artis zee MP 10094139 Artis Q floor 10848280 Artis Q ceiling 10848281 Artis Q biplane 10848282 Artis Q.zen floor 10848353 Artis Q.zen ceiling 10848354 Artis Q.zen biplane 10848355 Angiography systems developed for single and biplane diagnostic imaging. Reason: Due to a software error, the IAS (Image Acquisition System) may sporadically fail during startup/ restart, and result in delay of clinical treatment.

Recalling firm
Siemens Medical Solutions USA, Inc
Classification
Class II
Status
Terminated
Initiated
2021-06-30
Terminated
2023-02-10
Firm location
Malvern, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Artis Models with SW version VD12 listed below: Artis zee floor 10094135 Artis zee ceiling 10094137 Artis zee MP 10094139 Artis Q floor 10848280 Artis Q ceiling 10848281 Artis Q biplane 10848282 Artis Q.zen floor 10848353 Artis Q.zen ceiling 10848354 Artis Q.zen biplane 10848355 Angiography systems developed for single and biplane diagnostic imaging

Reason for recall

Due to a software error, the IAS (Image Acquisition System) may sporadically fail during startup/ restart, and result in delay of clinical treatment