Recall Z-2400-2020— Carefusion 2200 Inc

Class II · Terminated

Class II FDA medical device recall by Carefusion 2200 Inc, initiated 2020-05-18, terminated 2021-04-25. FDA's recall record names no specific 510(k) clearance for this event. Product: Mueller Knife Handle, Hexagonal - Product Usage: Intended to hold surgical blades.. Reason: Becton Dickinson and Company, is conducting a voluntary recall on the V. MuellerTM 3H Knife Handles catalog number listed in the table above. BD has been informed by our supplier that the external threads on some of the collets (Figure 1) contain a machining burr, potentially causing the collet to not fully seat in the knife handle (Figure 2), inhibiting complete grip of the inserted knife blade possibly resulting in a loose blade..

Recalling firm
Carefusion 2200 Inc
Classification
Class II
Status
Terminated
Initiated
2020-05-18
Terminated
2021-04-25
Firm location
Vernon Hills, IL, United States

Product description

Mueller Knife Handle, Hexagonal - Product Usage: Intended to hold surgical blades.

Reason for recall

Becton Dickinson and Company, is conducting a voluntary recall on the V. MuellerTM 3H Knife Handles catalog number listed in the table above. BD has been informed by our supplier that the external threads on some of the collets (Figure 1) contain a machining burr, potentially causing the collet to not fully seat in the knife handle (Figure 2), inhibiting complete grip of the inserted knife blade possibly resulting in a loose blade.