Recall Z-2399-2021— DePuy Orthopaedics, Inc.
Class II · Terminated
Class II FDA medical device recall by DePuy Orthopaedics, Inc., initiated 2021-08-10, terminated 2023-05-17. FDA's recall record names no specific 510(k) clearance for this event. Product: TRUMATCH CT CUT GUIDE KIT R- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting Part Number: 420916. Reason: Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to potentially result in limb malalignment..
- Recalling firm
- DePuy Orthopaedics, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2021-08-10
- Terminated
- 2023-05-17
- Firm location
- Warsaw, IN, United States
Product description
TRUMATCH CT CUT GUIDE KIT R- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting Part Number: 420916
Reason for recall
Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to potentially result in limb malalignment.
Data sourced from openFDA. This site is unofficial and independent of the FDA.