Recall Z-2397-2023— Philips North America
Class II · Ongoing
Class II FDA medical device recall by Philips North America, initiated 2023-06-12. It names one FDA 510(k) clearance: K213583. Product: MR 7700 Magnetic Resonance (MR) system, Product Numbers 782120 and 782130 (Upgrade to MR 7700). Reason: The system operator is able to bypass the SmokeDetector Interlock system locking mechanism after a smoke detection by power cycling the system..
- Recalling firm
- Philips North America
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-06-12
- Firm location
- Cambridge, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
MR 7700 Magnetic Resonance (MR) system, Product Numbers 782120 and 782130 (Upgrade to MR 7700)
Reason for recall
The system operator is able to bypass the SmokeDetector Interlock system locking mechanism after a smoke detection by power cycling the system.