Recall Z-2396-2021— DePuy Orthopaedics, Inc.

Class II · Terminated

Class II FDA medical device recall by DePuy Orthopaedics, Inc., initiated 2021-08-10, terminated 2023-05-17. FDA's recall record names no specific 510(k) clearance for this event. Product: TRUMATCH CT PIN GUIDE FEM R- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting Part Number: 420581. Reason: Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to potentially result in limb malalignment..

Recalling firm
DePuy Orthopaedics, Inc.
Classification
Class II
Status
Terminated
Initiated
2021-08-10
Terminated
2023-05-17
Firm location
Warsaw, IN, United States

Product description

TRUMATCH CT PIN GUIDE FEM R- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting Part Number: 420581

Reason for recall

Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to potentially result in limb malalignment.

Data sourced from openFDA. This site is unofficial and independent of the FDA.