Recall Z-2390-2018— Roche Diagnostics Corporation

Class II · Terminated

Class II FDA medical device recall by Roche Diagnostics Corporation, initiated 2018-04-27, terminated 2020-05-11. It names one FDA 510(k) clearance: K162840. Product: Elecsys Vitamin D total II, Vitamin D Test System, Material number 07464215160 Product Usage: This assay is intended for the quantitative determination of total 25-hydroxyvitamin D in human serum and plasma. This assay is to be used as an aid in the assessment of vitamin D sufficiency in adults. The electrochemiluminescence binding assay is intended for use on the immunoassay analyzer.. Reason: The device may give a falsely elevated result that is non-reproducible. If vitamin D level is monitored by laboratory tests and adapted based on the result, a dosage reduction of vitamin D supplementation cannot be excluded, and as a consequence, vitamin D deficiency may develop over time..

Recalling firm
Roche Diagnostics Corporation
Classification
Class II
Status
Terminated
Initiated
2018-04-27
Terminated
2020-05-11
Firm location
Indianapolis, IN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Elecsys Vitamin D total II, Vitamin D Test System, Material number 07464215160 Product Usage: This assay is intended for the quantitative determination of total 25-hydroxyvitamin D in human serum and plasma. This assay is to be used as an aid in the assessment of vitamin D sufficiency in adults. The electrochemiluminescence binding assay is intended for use on the immunoassay analyzer.

Reason for recall

The device may give a falsely elevated result that is non-reproducible. If vitamin D level is monitored by laboratory tests and adapted based on the result, a dosage reduction of vitamin D supplementation cannot be excluded, and as a consequence, vitamin D deficiency may develop over time.