Recall Z-2389-2016— Synthes (USA) Products LLC
Class II · Terminated
Class II FDA medical device recall by Synthes (USA) Products LLC, initiated 2016-06-28, terminated 2017-02-02. It names 5 FDA 510(k) clearances: K103243, K102694, K063049 and 2 more. Product: Synthes 2.4mm Titanium Locking Screws Self-Tapping with Stardrive Recess 12mm, Synthes 2.4mm Titanium Locking Screws Self-Tapping with Stardrive Recess 12mm, Plate, Fixation, Bone, Orthopedic Devices The 2.4mm Titanium Locking Screw Self-Tapping with Stardrive Recess 12mm is a part of the following systems: - 2.4. mm Titanium LCP Distal Radius Plate System - Titanium LCP Compact Distal Radius System - 2.4 mm Titanium LCP Radial Head Plate Set - Titanium Modular Mini Fragment LCP System - 2.4 mm Titanium Variable Angle LCP Distal Radius System - 2.7 mm/3.5 mm Titanium Distal Fibula Plate System. Reason: It was reported that the specified lots of the 2.4mm Titanium Locking Screw Self-Tapping with Stardrive Recess 12mm can break between the head and the shaft. This resulted from the neck feature being too close to the StarDrive feature leaving a thin wall between neck and drive..
- Recalling firm
- Synthes (USA) Products LLC
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-06-28
- Terminated
- 2017-02-02
- Firm location
- West Chester, PA, United States
510(k) clearances named in this recall
5 on FDA's record- K103243 — 2.4MM VA-LCP INTERCARPAL FUSION SYSTEM
- K102694 — 2.4MM VA-LCP DORSAL DISTAL RADIUS PLATES
- K063049 — SYNTHES (USA) MODULAR MINI FRAGMENT LCP SYSTEM
- K050110 — SYNTHES (USA) LCP MODULAR FOOT PLATES
- K012114 — SYNTHES LOCKING DISTAL RADIUS PLATING SYSTEM
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Synthes 2.4mm Titanium Locking Screws Self-Tapping with Stardrive Recess 12mm, Synthes 2.4mm Titanium Locking Screws Self-Tapping with Stardrive Recess 12mm, Plate, Fixation, Bone, Orthopedic Devices The 2.4mm Titanium Locking Screw Self-Tapping with Stardrive Recess 12mm is a part of the following systems: - 2.4. mm Titanium LCP Distal Radius Plate System - Titanium LCP Compact Distal Radius System - 2.4 mm Titanium LCP Radial Head Plate Set - Titanium Modular Mini Fragment LCP System - 2.4 mm Titanium Variable Angle LCP Distal Radius System - 2.7 mm/3.5 mm Titanium Distal Fibula Plate System
Reason for recall
It was reported that the specified lots of the 2.4mm Titanium Locking Screw Self-Tapping with Stardrive Recess 12mm can break between the head and the shaft. This resulted from the neck feature being too close to the StarDrive feature leaving a thin wall between neck and drive.