Recall Z-2387-2023— Covidien LP
Class I · Ongoing
Class I FDA medical device recall by Covidien LP, initiated 2023-06-28. It names 2 FDA 510(k) clearances: K102605, K020089. Product: MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 16 cm, Curved Extensions, Kit REF: 8888345611HP. Reason: Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus.
- Recalling firm
- Covidien LP
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2023-06-28
- Firm location
- Mansfield, MA, United States
510(k) clearances named in this recall
2 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 16 cm, Curved Extensions, Kit REF: 8888345611HP
Reason for recall
Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
Data sourced from openFDA. This site is unofficial and independent of the FDA.