Recall Z-2385-2026— ARROW INTERNATIONAL, LLC
Class I · Ongoing
Class I FDA medical device recall by ARROW INTERNATIONAL, LLC, initiated 2026-04-21. It names 4 FDA 510(k) clearances: K011761, K002507, K940079 and 1 more. Product: Description/REF: ACCESS TRAY/GU-04020; PSI KIT: 8.5 FR/AK-09803-CDC; PSI KIT: 9 FR/AK-09903-CDC; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-09810-PCMH1; PSI KIT: 9 FR X 10 CM/ASK-09903-KH; PSI KIT: 9 FR X 10 CM/ASK-09903-PHP1; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-09903-UPA; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09903-YNH; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09907-SL; PSI KIT: 8.5 FR X 4-1/8IN (10 CM)/NA-09803-CDC; PSI KIT: 9 FR X 4-1/8IN (10 CM)/NA-09903-CDC; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-09903-ECL4; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09903-MGH3; MULTI-LUMEN/PSI KIT/AK-11142-SPCS; MAC KIT: 2-L 9 FR DISTAL X 11.5 CM/CDC-11242-1A; MULTI-LUMEN/PSI KIT/AK-11142-D; MULTILUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-11142-MGH4; MULTI-LUMEN/PSI KIT/NA-11242-CDC; MULTI-LUMEN/PSI KIT/AK-21142-SK; MAC KIT: 2L 9 FR DISTAL X 10 CM AGB/ASK-21142-LMDTG; MULTI LUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-21142-RH; MULTILUMEN/PSI KIT 9 FR X 11.5 CM/ASK-21242-IMC; MAC KIT: 2L 9 FR DISTAL X 11.5 CM AGB/ASK-21242-LMDTG; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UMD; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UU3; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-VCU; MAC KIT: 2L 9 FR DISTAL X 11.5 CM AGB/CDC-21242-1A; MAC KIT: 2L 9 FR DISTAL X 11.5 CM AGB/CDC-21242-X1A; MULTI-LUMEN/PSI KIT/AK-21242-CDC; 2L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-FH4; MULTI LUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-21142-GH1; 2L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-SUU; 2-L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-UPMC4; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UKL; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UMF; MULTILUMEN/PSI KIT: 9 FR X 4 1/2 IN/ASK-21242-UR1; PSI KIT: 8.5 FR/AK-29803-CDC; PSI KIT: 9 FR/AK-29903-CDC; PSI KIT: 9 FR/AK-29907-CDC; PSI KIT: 8.5 FR X 4IN (10 CM)/ASK-09804-UN1; PSI KIT: 9 FR/ASK-09903-UM; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/ASK-29803-LMDT; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/ASK-29803-LMDTG; PSI KIT: 9 FR X 4 1/8 (10 CM)/ASK-29903-BID1; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-LMDTG; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-UMD; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-29903-UR1; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/CDC-29803-X1A; PSI KIT: 8.5 FR X 10 CM/ASK-29803-ECL; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-29803-UPM3; PSI KIT: 9 FR X 10 CM/ASK-29903-ECL; PSI KIT: 9FR X 4IN (10 CM) ANTIMICROBIAL/ASK-29903-UPM3; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-09907-TG1. Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries..
- Recalling firm
- ARROW INTERNATIONAL, LLC
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2026-04-21
- Firm location
- Morrisville, NC, United States
510(k) clearances named in this recall
4 on FDA's record- K011761 — MAC TWO-LUMEN CENTRAL VENOUS KIT WITH ARROWG ARD BLUE ANTIMICROBIAL SURFACE AND INTEGRAL HEMOSTASIS VALVE/SIDE PORT
- K002507 — TWO-LUMEN CENTRAL VENOUS ACCESS KIT WITH HEMOSTASIS VALVE/SIDE PORT
- K940079 — ARROW PERCUTANEOUS SHEATH INTRODUCER KIT AK-07903
- K780532 — PERCUTANEOUS SHEATH INTRODUCER KIT
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Description/REF: ACCESS TRAY/GU-04020; PSI KIT: 8.5 FR/AK-09803-CDC; PSI KIT: 9 FR/AK-09903-CDC; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-09810-PCMH1; PSI KIT: 9 FR X 10 CM/ASK-09903-KH; PSI KIT: 9 FR X 10 CM/ASK-09903-PHP1; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-09903-UPA; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09903-YNH; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09907-SL; PSI KIT: 8.5 FR X 4-1/8IN (10 CM)/NA-09803-CDC; PSI KIT: 9 FR X 4-1/8IN (10 CM)/NA-09903-CDC; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-09903-ECL4; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09903-MGH3; MULTI-LUMEN/PSI KIT/AK-11142-SPCS; MAC KIT: 2-L 9 FR DISTAL X 11.5 CM/CDC-11242-1A; MULTI-LUMEN/PSI KIT/AK-11142-D; MULTILUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-11142-MGH4; MULTI-LUMEN/PSI KIT/NA-11242-CDC; MULTI-LUMEN/PSI KIT/AK-21142-SK; MAC KIT: 2L 9 FR DISTAL X 10 CM AGB/ASK-21142-LMDTG; MULTI LUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-21142-RH; MULTILUMEN/PSI KIT 9 FR X 11.5 CM/ASK-21242-IMC; MAC KIT: 2L 9 FR DISTAL X 11.5 CM AGB/ASK-21242-LMDTG; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UMD; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UU3; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-VCU; MAC KIT: 2L 9 FR DISTAL X 11.5 CM AGB/CDC-21242-1A; MAC KIT: 2L 9 FR DISTAL X 11.5 CM AGB/CDC-21242-X1A; MULTI-LUMEN/PSI KIT/AK-21242-CDC; 2L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-FH4; MULTI LUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-21142-GH1; 2L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-SUU; 2-L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-UPMC4; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UKL; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UMF; MULTILUMEN/PSI KIT: 9 FR X 4 1/2 IN/ASK-21242-UR1; PSI KIT: 8.5 FR/AK-29803-CDC; PSI KIT: 9 FR/AK-29903-CDC; PSI KIT: 9 FR/AK-29907-CDC; PSI KIT: 8.5 FR X 4IN (10 CM)/ASK-09804-UN1; PSI KIT: 9 FR/ASK-09903-UM; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/ASK-29803-LMDT; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/ASK-29803-LMDTG; PSI KIT: 9 FR X 4 1/8 (10 CM)/ASK-29903-BID1; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-LMDTG; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-UMD; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-29903-UR1; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/CDC-29803-X1A; PSI KIT: 8.5 FR X 10 CM/ASK-29803-ECL; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-29803-UPM3; PSI KIT: 9 FR X 10 CM/ASK-29903-ECL; PSI KIT: 9FR X 4IN (10 CM) ANTIMICROBIAL/ASK-29903-UPM3; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-09907-TG1
Reason for recall
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.