Recall Z-2384-2012— Philips Healthcare Inc.
Class II · Terminated
Class II FDA medical device recall by Philips Healthcare Inc., initiated 2012-08-14, terminated 2017-01-17. It names one FDA 510(k) clearance: K982795. Product: Philips DigitalDiagnost Systems with M Cabinet Model: 712025 Stationary radiographic system, digital-Universal RAD Applications. Reason: Creepage Distance between the two primary phases in the GEO Transformer is 4.5 mm.does not meet (IEC) International standard of 5.5 mm..
- Recalling firm
- Philips Healthcare Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2012-08-14
- Terminated
- 2017-01-17
- Firm location
- Andover, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Philips DigitalDiagnost Systems with M Cabinet Model: 712025 Stationary radiographic system, digital-Universal RAD Applications
Reason for recall
Creepage Distance between the two primary phases in the GEO Transformer is 4.5 mm.does not meet (IEC) International standard of 5.5 mm.