Recall Z-2375-2026— ARROW INTERNATIONAL, LLC

Class I · Ongoing

Class I FDA medical device recall by ARROW INTERNATIONAL, LLC, initiated 2026-04-21. FDA's recall record names no specific 510(k) clearance for this event. Product: Description/REF: PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PNHS2; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-PNHS2; PI CVC KIT: 4L 8.5 FR X 20 CM AGB+/ASK-45854-PNHS2. Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries..

Recalling firm
ARROW INTERNATIONAL, LLC
Classification
Class I
Status
Ongoing
Initiated
2026-04-21
Firm location
Morrisville, NC, United States

Product description

Description/REF: PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PNHS2; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-PNHS2; PI CVC KIT: 4L 8.5 FR X 20 CM AGB+/ASK-45854-PNHS2

Reason for recall

Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.