Recall Z-2369-2026— ARROW INTERNATIONAL, LLC
Class I · Ongoing
Class I FDA medical device recall by ARROW INTERNATIONAL, LLC, initiated 2026-04-21. FDA's recall record names no specific 510(k) clearance for this event. Product: Description/REF: PSI KIT: 7 FR X 4IN ANTIMICROBIAL/ASK-09701-UMH; PSI KIT: 8.5 FR X 4 1/8 IN (10 CM)/ASK-09803-OSU4; PSI/CVC KIT: 9 FR/7 FR X 6IN (16CM)/ASK-09903-AMC1; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09903-CMC1; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09903-OSU4; PSI KIT: 9 FR X 4 1/8 IN (10CM)/ASK-09903-PCMH1; PSI KIT: 8.5 FR X 10 CM/CDC-09803-1A; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-09903-ECL3; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-11242-PHP2; MULTI LUMEN/PSI KIT/ASK-11142-NSU; MULTI-LUMEN/PSI KIT/AK-21242-X; MAC KIT: 2L 9 FR DISTAL X 10 CM AGB/ASK-21142-DMC; MAC KIT: 2L 9 FR DISTAL X 10 CM AGB/ASK-21142-GMC1; MULTILUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-21142-HF5; MAC KIT: 2L 9 FR DISTAL X 10 CM AGB/ASK-21142-LMDT; 2L MAC KIT: 9 FR DISTAL X 4IN (10 CM)/ASK-21142-UCL1; 2-L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-UPMC2; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-SL1; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UMH; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UN1; MAC KIT: 2-L 9 FR DISTAL X 10 CM AGB/CDC-21142-XCN1A; MAC KIT: 2-L 9 FR DISTAL X 10 CM AGB/CDC-21142-XN1A; MAC KIT: 2-L 9 FR DISTAL X 11.5 CM AGB/CDC-21242-XCN1A; MAC KIT: 2-L 9 FR DISTAL X 11.5 CM AGB/CDC-21242-XN1A; 2L MAC KIT: 9 FR DISTAL X 4 IN/ASK-21142-CO1; 2L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-FH3; 2L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-OHSU3; MULTILUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-21142-SG2; MULTILUMEN/MAC KIT: 9 FR X 4 IN (10 CM)/ASK-21142-UOI1; 2 LUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-21142-WHC2; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-BID1; AGB MAC KIT: 2L 9 FR DISTAL/ASK-21242-BSWH2; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-DU1; MULTILUMEN/PSI KIT 9 FR X 11.5 CM/ASK-21242-PCMH1; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UCH2; AGB MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UHC2; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/ASK-29803-BSWH2; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-29803-HHC; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-29803-MCY1; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-29803-UPM1; PSI KIT: 8.5 FR X 4 1/8 IN (10 CM)/ASK-29804-VCU; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-MCY1; PSI KIT: 9 FR/ASK-29903-MHI; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-29903-OHU4; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-29903-UCH2; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-UCL1; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-UHC4; PSI KIT: 9FR X 4IN (10 CM) ANTIMICROBIAL/ASK-29903-UPM1; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-29907-LGH; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/CDC-29803-1A; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/CDC-29903-1A; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/CDC-29903-X1A; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/CDC-29903-XCN1A; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/CDC-29903-XN1A; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-AHH; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/CDC-29803-XN1A; PSI ACCESS TRAY/ASK-04001-MM2; MAC KIT: 2L 9 FR DISTAL X 10 CM AGB/ASK-21142-GMC; MULTI LUMEN/PSI KIT: 9 FR DISTAL X 4 1/2/ASK-21242-OSU3. Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries..
- Recalling firm
- ARROW INTERNATIONAL, LLC
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2026-04-21
- Firm location
- Morrisville, NC, United States
Product description
Description/REF: PSI KIT: 7 FR X 4IN ANTIMICROBIAL/ASK-09701-UMH; PSI KIT: 8.5 FR X 4 1/8 IN (10 CM)/ASK-09803-OSU4; PSI/CVC KIT: 9 FR/7 FR X 6IN (16CM)/ASK-09903-AMC1; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09903-CMC1; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09903-OSU4; PSI KIT: 9 FR X 4 1/8 IN (10CM)/ASK-09903-PCMH1; PSI KIT: 8.5 FR X 10 CM/CDC-09803-1A; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-09903-ECL3; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-11242-PHP2; MULTI LUMEN/PSI KIT/ASK-11142-NSU; MULTI-LUMEN/PSI KIT/AK-21242-X; MAC KIT: 2L 9 FR DISTAL X 10 CM AGB/ASK-21142-DMC; MAC KIT: 2L 9 FR DISTAL X 10 CM AGB/ASK-21142-GMC1; MULTILUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-21142-HF5; MAC KIT: 2L 9 FR DISTAL X 10 CM AGB/ASK-21142-LMDT; 2L MAC KIT: 9 FR DISTAL X 4IN (10 CM)/ASK-21142-UCL1; 2-L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-UPMC2; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-SL1; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UMH; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UN1; MAC KIT: 2-L 9 FR DISTAL X 10 CM AGB/CDC-21142-XCN1A; MAC KIT: 2-L 9 FR DISTAL X 10 CM AGB/CDC-21142-XN1A; MAC KIT: 2-L 9 FR DISTAL X 11.5 CM AGB/CDC-21242-XCN1A; MAC KIT: 2-L 9 FR DISTAL X 11.5 CM AGB/CDC-21242-XN1A; 2L MAC KIT: 9 FR DISTAL X 4 IN/ASK-21142-CO1; 2L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-FH3; 2L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-OHSU3; MULTILUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-21142-SG2; MULTILUMEN/MAC KIT: 9 FR X 4 IN (10 CM)/ASK-21142-UOI1; 2 LUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-21142-WHC2; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-BID1; AGB MAC KIT: 2L 9 FR DISTAL/ASK-21242-BSWH2; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-DU1; MULTILUMEN/PSI KIT 9 FR X 11.5 CM/ASK-21242-PCMH1; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UCH2; AGB MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UHC2; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/ASK-29803-BSWH2; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-29803-HHC; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-29803-MCY1; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-29803-UPM1; PSI KIT: 8.5 FR X 4 1/8 IN (10 CM)/ASK-29804-VCU; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-MCY1; PSI KIT: 9 FR/ASK-29903-MHI; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-29903-OHU4; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-29903-UCH2; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-UCL1; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-UHC4; PSI KIT: 9FR X 4IN (10 CM) ANTIMICROBIAL/ASK-29903-UPM1; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-29907-LGH; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/CDC-29803-1A; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/CDC-29903-1A; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/CDC-29903-X1A; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/CDC-29903-XCN1A; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/CDC-29903-XN1A; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-AHH; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/CDC-29803-XN1A; PSI ACCESS TRAY/ASK-04001-MM2; MAC KIT: 2L 9 FR DISTAL X 10 CM AGB/ASK-21142-GMC; MULTI LUMEN/PSI KIT: 9 FR DISTAL X 4 1/2/ASK-21242-OSU3
Reason for recall
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
Data sourced from openFDA. This site is unofficial and independent of the FDA.