Recall Z-2367-2019— Siemens Medical Solutions USA, Inc

Class II · Terminated

Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2019-07-30, terminated 2020-04-29. It names one FDA 510(k) clearance: K181407. Product: Artis zee and Artis Q/ Q.zen Interventional Fluoroscopic X-Ray System with patient tables as follows: Systems Material Number Artis zee biplane 10094141; Artis Q floor 10848280; Artis Q biplane 10848282; Artis Q.zen floor 10848353; Artis Q.zen biplane 10848355; Artis zee ceiling 10094137; Artis Q ceiling 10848281; Artis Q zeego 10848283 ArtisQ.zen ceiling 10848354 Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. The Artis family can also support the acquisition of position triggered imaging for spatial data synthesis.. Reason: Artis zee/Q systems patient tables may potentially have cracks in the table mainframe, and cause mechanical detachment of the upper part of the patient bed and lead to hazardous situations for patients and medical personnel due to falling parts.

Recalling firm
Siemens Medical Solutions USA, Inc
Classification
Class II
Status
Terminated
Initiated
2019-07-30
Terminated
2020-04-29
Firm location
Malvern, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Artis zee and Artis Q/ Q.zen Interventional Fluoroscopic X-Ray System with patient tables as follows: Systems Material Number Artis zee biplane 10094141; Artis Q floor 10848280; Artis Q biplane 10848282; Artis Q.zen floor 10848353; Artis Q.zen biplane 10848355; Artis zee ceiling 10094137; Artis Q ceiling 10848281; Artis Q zeego 10848283 ArtisQ.zen ceiling 10848354 Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. The Artis family can also support the acquisition of position triggered imaging for spatial data synthesis.

Reason for recall

Artis zee/Q systems patient tables may potentially have cracks in the table mainframe, and cause mechanical detachment of the upper part of the patient bed and lead to hazardous situations for patients and medical personnel due to falling parts