Recall Z-2366-2020— Ortho-Clinical Diagnostics

Class II · Terminated

Class II FDA medical device recall by Ortho-Clinical Diagnostics, initiated 2020-05-11, terminated 2021-09-22. It names one FDA 510(k) clearance: K812027. Product: VITROS Chemistry Product Performance Verifier I, Product Code 8231474, UDI # 10758750004577 - Product Usage: For in vitro diagnostic use only to quantitatively measure amylase activity in serum, plasma, and urine using VITROS 250/350/5, 1 FS/4600/XT 3400 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems.. Reason: Multiple coatings of VITROS Chemistry Products AMYL Slides are showing imprecision on control fluids when compared to the expected within lab SD as listed on the Performance Verifier (PV) Assay Sheets..

Recalling firm
Ortho-Clinical Diagnostics
Classification
Class II
Status
Terminated
Initiated
2020-05-11
Terminated
2021-09-22
Firm location
Rochester, NY, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

VITROS Chemistry Product Performance Verifier I, Product Code 8231474, UDI # 10758750004577 - Product Usage: For in vitro diagnostic use only to quantitatively measure amylase activity in serum, plasma, and urine using VITROS 250/350/5, 1 FS/4600/XT 3400 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems.

Reason for recall

Multiple coatings of VITROS Chemistry Products AMYL Slides are showing imprecision on control fluids when compared to the expected within lab SD as listed on the Performance Verifier (PV) Assay Sheets.

Data sourced from openFDA. This site is unofficial and independent of the FDA.