Recall Z-2362-2018— Zimmer Biomet, Inc.
Class II · Terminated
Class II FDA medical device recall by Zimmer Biomet, Inc., initiated 2018-05-30, terminated 2020-05-28. FDA's recall record names no specific 510(k) clearance for this event. Product: Zimmer¿ M/L Taper Hip Prosthesis Plasma Sprayed, Size 4 Standard Offset; 00-7711-004-10. Reason: M/L Taper Hip Prosthesis and M/L Taper Hip Prosthesis with Kinectiv Technology manufactured prior to October 2015 are being recalled due to the potential for black residue to be observed within the polyethylene bag containing the implant. The black residue is a combination of the titanium plasma spray used to create the porous surface of the implant and polyethylene bag containing the implant. The recalling firm states the residue is biocompatible. Risks include delay of surgery while locating a replacement device, and surgical intervention due to wear caused by debris..
- Recalling firm
- Zimmer Biomet, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-05-30
- Terminated
- 2020-05-28
- Firm location
- Warsaw, IN, United States
Product description
Zimmer¿ M/L Taper Hip Prosthesis Plasma Sprayed, Size 4 Standard Offset; 00-7711-004-10
Reason for recall
M/L Taper Hip Prosthesis and M/L Taper Hip Prosthesis with Kinectiv Technology manufactured prior to October 2015 are being recalled due to the potential for black residue to be observed within the polyethylene bag containing the implant. The black residue is a combination of the titanium plasma spray used to create the porous surface of the implant and polyethylene bag containing the implant. The recalling firm states the residue is biocompatible. Risks include delay of surgery while locating a replacement device, and surgical intervention due to wear caused by debris.
Data sourced from openFDA. This site is unofficial and independent of the FDA.