Recall Z-2360-2023— Siemens Medical Solutions USA, Inc

Class II · Ongoing

Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2023-06-19. It names one FDA 510(k) clearance: K220432. Product: ARTIS icono floor-Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Material Number: 11327700. Reason: Due to a hardware issue in the cable connectors of the system generator, a thermal overload in the cable connections may occur when performing excessive fluoroscopy/acquisitions, a burning smell may be detectable coming from the generator cabinet and the system may lose the imaging functionality of the corresponding plane and result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system.

Recalling firm
Siemens Medical Solutions USA, Inc
Classification
Class II
Status
Ongoing
Initiated
2023-06-19
Firm location
Malvern, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ARTIS icono floor-Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Material Number: 11327700

Reason for recall

Due to a hardware issue in the cable connectors of the system generator, a thermal overload in the cable connections may occur when performing excessive fluoroscopy/acquisitions, a burning smell may be detectable coming from the generator cabinet and the system may lose the imaging functionality of the corresponding plane and result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system