Recall Z-2359-2023— Siemens Medical Solutions USA, Inc
Class II · Ongoing
Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2023-06-19. It names one FDA 510(k) clearance: K220432. Product: ARTIS icono biplane -Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Material Number: 11327600. Reason: Due to a hardware issue in the cable connectors of the system generator, a thermal overload in the cable connections may occur when performing excessive fluoroscopy/acquisitions, a burning smell may be detectable coming from the generator cabinet and the system may lose the imaging functionality of the corresponding plane and result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system.
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-06-19
- Firm location
- Malvern, PA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
ARTIS icono biplane -Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Material Number: 11327600
Reason for recall
Due to a hardware issue in the cable connectors of the system generator, a thermal overload in the cable connections may occur when performing excessive fluoroscopy/acquisitions, a burning smell may be detectable coming from the generator cabinet and the system may lose the imaging functionality of the corresponding plane and result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system
Data sourced from openFDA. This site is unofficial and independent of the FDA.