Recall Z-2359-2019— Agilent Technologies, Inc.

Class II · Terminated

Class II FDA medical device recall by Agilent Technologies, Inc., initiated 2018-08-15, terminated 2021-09-28. FDA's recall record names no specific 510(k) clearance for this event. Product: MassHunter Quantitative Analysis Software, Version B.07.01, used with Triple Quadrupole Mass Spectrometers, models K6460 and K6420 Product Usage: A mass spectrometer for clinical use is a device intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass.. Reason: When utilizing specific software versions of an LC/MS device, and Batch at a Glance in Compound Table view, a defect occurs when a new sample(s) is inserted. This defect creates a mismatch between the sample name and the column header whereby the sample header will be offset by one sample. The recalling firm requests that you discontinue using this workflow to generate reports because the Quant batch table will not display fully analyzed results or save any changes made..

Recalling firm
Agilent Technologies, Inc.
Classification
Class II
Status
Terminated
Initiated
2018-08-15
Terminated
2021-09-28
Firm location
Santa Clara, CA, United States

Product description

MassHunter Quantitative Analysis Software, Version B.07.01, used with Triple Quadrupole Mass Spectrometers, models K6460 and K6420 Product Usage: A mass spectrometer for clinical use is a device intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass.

Reason for recall

When utilizing specific software versions of an LC/MS device, and Batch at a Glance in Compound Table view, a defect occurs when a new sample(s) is inserted. This defect creates a mismatch between the sample name and the column header whereby the sample header will be offset by one sample. The recalling firm requests that you discontinue using this workflow to generate reports because the Quant batch table will not display fully analyzed results or save any changes made.

Data sourced from openFDA. This site is unofficial and independent of the FDA.