Recall Z-2358-2020— Becton, Dickinson and Company, BD Biosciences
Class II · Terminated
Class II FDA medical device recall by Becton, Dickinson and Company, BD Biosciences, initiated 2019-07-08, terminated 2023-01-31. It names one FDA 510(k) clearance: K131301. Product: BD FACS Sample Prep Assistant (SPA) III - Product Usage: Prep Assistant III is intended to prepare human whole blood for flow cytometric analysis on BD FACSCanto II and BD FACSCalibur flow cytometry systems. Pipetting blood, reagents, and lysing solution using the following previously cleared assays for flow cytometric analysis on BD FACSCanto II flow cytometry systems. Reason: The recalling firm confirmed that SPA II and III users may encounter probes with unacceptable wear prior to reaching the 2500 piercings guideline indicated in the Instructions for Use (IFU). When the probe is worn, particles from the rubber cap are more likely to clog the probe, resulting in blocked or inadequate sample dispense..
- Recalling firm
- Becton, Dickinson and Company, BD Biosciences
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-07-08
- Terminated
- 2023-01-31
- Firm location
- San Jose, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
BD FACS Sample Prep Assistant (SPA) III - Product Usage: Prep Assistant III is intended to prepare human whole blood for flow cytometric analysis on BD FACSCanto II and BD FACSCalibur flow cytometry systems. Pipetting blood, reagents, and lysing solution using the following previously cleared assays for flow cytometric analysis on BD FACSCanto II flow cytometry systems
Reason for recall
The recalling firm confirmed that SPA II and III users may encounter probes with unacceptable wear prior to reaching the 2500 piercings guideline indicated in the Instructions for Use (IFU). When the probe is worn, particles from the rubber cap are more likely to clog the probe, resulting in blocked or inadequate sample dispense.
Data sourced from openFDA. This site is unofficial and independent of the FDA.