Recall Z-2354-2018— FHC, Inc.
Class II · Terminated
Class II FDA medical device recall by FHC, Inc., initiated 2018-05-10, terminated 2019-04-30. It names one FDA 510(k) clearance: K092562. Product: microTargeting" Lead Adapter for STar" Drive, Product Number 70-CN-DB. Reason: The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure..
- Recalling firm
- FHC, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-05-10
- Terminated
- 2019-04-30
- Firm location
- Bowdoin, ME, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
microTargeting" Lead Adapter for STar" Drive, Product Number 70-CN-DB
Reason for recall
The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.
Data sourced from openFDA. This site is unofficial and independent of the FDA.