Recall Z-2348-2012— Terumo Cardiovascular Systems Corporation

Class II · Terminated

Class II FDA medical device recall by Terumo Cardiovascular Systems Corporation, initiated 2009-08-12, terminated 2013-01-29. It names one FDA 510(k) clearance: K915183. Product: Safety monitor included with System 8000 Base The Sarns" Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment. The safety monitor is indicated to warn when the blood level in the reservoir reaches the operator-set limit, and to stop the arterial and cardioplegia pumps.. Reason: Terumo Cardiovascular Systems (Terumo CVS) received complaints from our Affiliate, Terumo Dubai that the safety monitors were tripping off. Terumo CVS' investigation determined that 8 monitors were manufactured with an underrated circuit breaker due to an error where an associate pulled an underrated circuit breaker from an incorrect inventory location. Since this time, inventory locations hav.

Recalling firm
Terumo Cardiovascular Systems Corporation
Classification
Class II
Status
Terminated
Initiated
2009-08-12
Terminated
2013-01-29
Firm location
Ann Arbor, MI, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Safety monitor included with System 8000 Base The Sarns" Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment. The safety monitor is indicated to warn when the blood level in the reservoir reaches the operator-set limit, and to stop the arterial and cardioplegia pumps.

Reason for recall

Terumo Cardiovascular Systems (Terumo CVS) received complaints from our Affiliate, Terumo Dubai that the safety monitors were tripping off. Terumo CVS' investigation determined that 8 monitors were manufactured with an underrated circuit breaker due to an error where an associate pulled an underrated circuit breaker from an incorrect inventory location. Since this time, inventory locations hav

Data sourced from openFDA. This site is unofficial and independent of the FDA.