Recall Z-2342-2012— Medtronic Xomed, Inc.

Class II · Terminated

Class II FDA medical device recall by Medtronic Xomed, Inc., initiated 2010-09-27, terminated 2013-11-26. It names one FDA 510(k) clearance: K083124. Product: Interfaces subject to recall: Medtronic NIM (Nerve Integrity Monitor) 3.0 Patient Interfaces 8253200 (Response) LABELING: Box Label: REF 8253200 x1***NIM-Response 3.0 Patient Interface***Medtronic Xomed, Inc., 6743 Southpoint Drive North Jacksonville, FL 32216-0980 USA***Rx Only***Lot***SN:***" Top of unit is labeled in part: "***NIM-Response 3.0***Medtronic***" Bottom/Underside of unit is labeled in part: "***5X20mm FA.L., 32mA. 250VAC***REF***MEDTRONIC XOMED INC. JACKSONVILLE, FL USA***" Cord is labeled in part: "***Ref.***Mfg Date***Sn***" The NIM 3.0 Patient Interfaces [REF #'s 8253210 (Pulse), 8253200 (Response), 8253410 (Neuro)] is an EMG Nerve Integrity Monitor intended for locating and identifying cranial and peripheral motor and mixed motor-sensory nerves during surgery, including spinal cord and spinal nerve roots.. Reason: In January 2010, Medtronic issued a recall of Medtronic NIM 3.0 Patient Interfaces after Medtronic identified a trend of open channel conditions (signal loss from the electrode). When the NIM 3.0 detects the signal loss from the electrode (i.e. open channel), it fails the electrode check and provides an audible and visual warning to the user which could also create a delay in the surgical procedu.

Recalling firm
Medtronic Xomed, Inc.
Classification
Class II
Status
Terminated
Initiated
2010-09-27
Terminated
2013-11-26
Firm location
Jacksonville, FL, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Interfaces subject to recall: Medtronic NIM (Nerve Integrity Monitor) 3.0 Patient Interfaces 8253200 (Response) LABELING: Box Label: REF 8253200 x1***NIM-Response 3.0 Patient Interface***Medtronic Xomed, Inc., 6743 Southpoint Drive North Jacksonville, FL 32216-0980 USA***Rx Only***Lot***SN:***" Top of unit is labeled in part: "***NIM-Response 3.0***Medtronic***" Bottom/Underside of unit is labeled in part: "***5X20mm FA.L., 32mA. 250VAC***REF***MEDTRONIC XOMED INC. JACKSONVILLE, FL USA***" Cord is labeled in part: "***Ref.***Mfg Date***Sn***" The NIM 3.0 Patient Interfaces [REF #'s 8253210 (Pulse), 8253200 (Response), 8253410 (Neuro)] is an EMG Nerve Integrity Monitor intended for locating and identifying cranial and peripheral motor and mixed motor-sensory nerves during surgery, including spinal cord and spinal nerve roots.

Reason for recall

In January 2010, Medtronic issued a recall of Medtronic NIM 3.0 Patient Interfaces after Medtronic identified a trend of open channel conditions (signal loss from the electrode). When the NIM 3.0 detects the signal loss from the electrode (i.e. open channel), it fails the electrode check and provides an audible and visual warning to the user which could also create a delay in the surgical procedu

Data sourced from openFDA. This site is unofficial and independent of the FDA.