Recall Z-2334-2025— Baxter Healthcare Corporation

Class I · Ongoing

Class I FDA medical device recall by Baxter Healthcare Corporation, initiated 2025-08-04. FDA's recall record names no specific 510(k) clearance for this event. Product: Novum IQ Syringe Pump, Product Code 40700BAX, infusion pump. Reason: Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the Novum IQ Large Volume Pump (LVP) and Novum IQ Syringe Pump (SP) due to software anomalies that may result in a blank Run screen (LVP and SP) and/or false motor movement system error (SP only)..

Recalling firm
Baxter Healthcare Corporation
Classification
Class I
Status
Ongoing
Initiated
2025-08-04
Firm location
Deerfield, IL, United States

Product description

Novum IQ Syringe Pump, Product Code 40700BAX, infusion pump

Reason for recall

Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the Novum IQ Large Volume Pump (LVP) and Novum IQ Syringe Pump (SP) due to software anomalies that may result in a blank Run screen (LVP and SP) and/or false motor movement system error (SP only).

Data sourced from openFDA. This site is unofficial and independent of the FDA.