Recall Z-2334-2018— St Jude Medical Inc.

Class II · Terminated

Class II FDA medical device recall by St Jude Medical Inc., initiated 2018-04-16, terminated 2021-02-04. It names one FDA 510(k) clearance: K163407. Product: St. Jude Medical CONFIRM Rx Insertable Cardiac Monitor, Sterile EO, Model #: DM3500 The Confirm Rx ICM is indicated for the monitoring and diagnostic evaluation of patients who experience unexplained symptoms such as: dizziness, palpitations, chest pain, syncope, and shortness of breath, as well as patients who are at risk for cardiac arrhythmias. It is also indicated for patients who have been previously diagnosed with atrial fibrillation or who are susceptible to developing atrial fibrillation.. Reason: The firm is advising physicians that exposure to sub-freezing temperatures during the supply chain process caused a transient battery voltage drop for a small number of Confirm Rx Model DM3500 Insertable Cardiac Monitoring (ICM) devices. This drop is an expected and normal behavior for this battery chemistry, but causes an incorrect display of a low battery indicator even after the battery voltage returns to normal..

Recalling firm
St Jude Medical Inc.
Classification
Class II
Status
Terminated
Initiated
2018-04-16
Terminated
2021-02-04
Firm location
Sylmar, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

St. Jude Medical CONFIRM Rx Insertable Cardiac Monitor, Sterile EO, Model #: DM3500 The Confirm Rx ICM is indicated for the monitoring and diagnostic evaluation of patients who experience unexplained symptoms such as: dizziness, palpitations, chest pain, syncope, and shortness of breath, as well as patients who are at risk for cardiac arrhythmias. It is also indicated for patients who have been previously diagnosed with atrial fibrillation or who are susceptible to developing atrial fibrillation.

Reason for recall

The firm is advising physicians that exposure to sub-freezing temperatures during the supply chain process caused a transient battery voltage drop for a small number of Confirm Rx Model DM3500 Insertable Cardiac Monitoring (ICM) devices. This drop is an expected and normal behavior for this battery chemistry, but causes an incorrect display of a low battery indicator even after the battery voltage returns to normal.