Recall Z-2328-2016— Philips Electronics North America Corporation
Class II · Terminated
Class II FDA medical device recall by Philips Electronics North America Corporation, initiated 2016-07-26, terminated 2020-07-14. It names 3 FDA 510(k) clearances: K161531, K141015, K131872. Product: IntelliVue Patient Monitor Product Usage: The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The monitors are intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitors are intended for use by trained healthcare professionals in a hospital environment.. Reason: If an affected Patient Monitor has been powered on continuously for several months, any displayed waveforms will contain outdated data and therefore fail to reflect the patients current condition..
- Recalling firm
- Philips Electronics North America Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-07-26
- Terminated
- 2020-07-14
- Firm location
- Andover, MA, United States
510(k) clearances named in this recall
3 on FDA's record- K161531 — IntelliVue Patient Monitors MP2, MP5, MP5T, MP5SC, and Multi-Measurement Module X2 +4 more
- K141015 — INTELLIVUE PATIENT MONITOR MX400, INTELLIVUE PATIENT MONITOR MX450, INTELLIVUE PATIENT MONITOR MX500, INTELLIVUE PATIENT
- K131872 — INTELLIVUE PATIENT MONITOR MX500, INTELLIVUE PATIENT MONITOR MX550
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
IntelliVue Patient Monitor Product Usage: The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The monitors are intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitors are intended for use by trained healthcare professionals in a hospital environment.
Reason for recall
If an affected Patient Monitor has been powered on continuously for several months, any displayed waveforms will contain outdated data and therefore fail to reflect the patients current condition.