Recall Z-2325-2018— St Jude Medical Inc.
Class II · Terminated
Class II FDA medical device recall by St Jude Medical Inc., initiated 2018-04-16, terminated 2024-03-08. FDA's recall record names no specific 510(k) clearance for this event. Product: Quadra Assura MP, Sterile EO, Model #/ Part #: CD3269-40/100043177; CD3269-40Q/100043178; CD3271-40/100043130; CD3271-40Q/100043191; CD3369-40C/100080339, 100080478; CD3369-40Q/100080439, 100080476, 100105528; CD3371-40/100078814, 100079232, 100079513; CD3371-40C/100078639, 100079231, 100079478, 100105493, 100105494; CD3371-40Q/100078815,100079282, 100079479, 100138237; CD3371-40QC/100078640, 100079165, 100079506, 100079541, 100105505, 100105506, 100138173, 100138179, 100138222, 100138234; CD3387-40C/100105525; CD3387-40QC/100138158. Reason: The firm is releasing new firmware for high voltage implantable cardiac devices (ICD) and cardiac resynchronization therapy defibrillators (CRT-D) intended to strengthen security by providing an additional layer of protection against unauthorized device access. For older generation devices not capable of accepting the firmware, the firm is advising customers of the option to permanently disable the RF communication capability in these devices..
- Recalling firm
- St Jude Medical Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-04-16
- Terminated
- 2024-03-08
- Firm location
- Sylmar, CA, United States
Product description
Quadra Assura MP, Sterile EO, Model #/ Part #: CD3269-40/100043177; CD3269-40Q/100043178; CD3271-40/100043130; CD3271-40Q/100043191; CD3369-40C/100080339, 100080478; CD3369-40Q/100080439, 100080476, 100105528; CD3371-40/100078814, 100079232, 100079513; CD3371-40C/100078639, 100079231, 100079478, 100105493, 100105494; CD3371-40Q/100078815,100079282, 100079479, 100138237; CD3371-40QC/100078640, 100079165, 100079506, 100079541, 100105505, 100105506, 100138173, 100138179, 100138222, 100138234; CD3387-40C/100105525; CD3387-40QC/100138158
Reason for recall
The firm is releasing new firmware for high voltage implantable cardiac devices (ICD) and cardiac resynchronization therapy defibrillators (CRT-D) intended to strengthen security by providing an additional layer of protection against unauthorized device access. For older generation devices not capable of accepting the firmware, the firm is advising customers of the option to permanently disable the RF communication capability in these devices.
Data sourced from openFDA. This site is unofficial and independent of the FDA.