Recall Z-2325-2016— B-K Medical A/S

Class II · Terminated

Class II FDA medical device recall by B-K Medical A/S, initiated 2016-06-28, terminated 2016-08-29. It names 4 FDA 510(k) clearances: K151910, K152052, K143298 and 1 more. Product: BK Medical ApS Ultrasound System Scanner bk2300 - Model #: bk3500 with software releases 1.2.0, 1.2.1 and 1.2.2. The system is intended for diagnostic ultrasound imaging or fluid flow analysis of the human body, data processing and guidance of puncture and biopsy. The system performs simple geometric measurements and calculations in the following areas: Emergency Medicine, Anesthesia, MSK, Vascular, Cardiology, OB/GYN. Reason: In the system (bk3500) software generic volume measurement (HxWxLxl) function the user may be unaware of the calculation factor (1) and assume the generic volume calculation factor is similar to that of the other (prostate, adenoma, testis, kidney, and bladder) volume calculation factors provided in the software..

Recalling firm
B-K Medical A/S
Classification
Class II
Status
Terminated
Initiated
2016-06-28
Terminated
2016-08-29
Firm location
Herlev, Denmark

510(k) clearances named in this recall

4 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

BK Medical ApS Ultrasound System Scanner bk2300 - Model #: bk3500 with software releases 1.2.0, 1.2.1 and 1.2.2. The system is intended for diagnostic ultrasound imaging or fluid flow analysis of the human body, data processing and guidance of puncture and biopsy. The system performs simple geometric measurements and calculations in the following areas: Emergency Medicine, Anesthesia, MSK, Vascular, Cardiology, OB/GYN

Reason for recall

In the system (bk3500) software generic volume measurement (HxWxLxl) function the user may be unaware of the calculation factor (1) and assume the generic volume calculation factor is similar to that of the other (prostate, adenoma, testis, kidney, and bladder) volume calculation factors provided in the software.