Recall Z-2324-2024— Zoll Medical Corporation

Class II · Ongoing

Class II FDA medical device recall by Zoll Medical Corporation, initiated 2024-05-31. FDA's recall record names no specific 510(k) clearance for this event. Product: ZOLL Powerheart G5 AED, Semi-Automatic, G5Sxxx Family -Automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). Model/Catalog Number: G5S-00A G5S-00A-TSO G5S-00C G5S-01A G5S-02A G5S-02C G5S-02-L G5S-04A G5S-05A G5S-05C G5S-06A G5S-06C G5S-08A G5S-08C G5S-10A G5S-10C G5S-11A G5S-11C G5S-12C G5S-13A G5S-14A G5S-15A G5S-17A G5S-17C G5S-19A G5S-23C G5S-29A G5S-29C G5S-31A G5S-31A-SJA G5S-31C-SJA G5S-36A G5S-37A G5S-41A G5S-41C G5S-80A G5S-80A-TSO G5S-80C G5S-80-L G5S-82A G5S-82C G5S-83C G5S-90C. Reason: G5 Semi-Automatic AED is shipped, with a protective film over its front panel, to protect the screen and shock button from cosmetic damage during shipping. It has come to our attention that customers may not be removing the protective film during deployment of the product. If the protective film is left adhered to the front bezel it may prevent the user from actuating the shock button and may lead to a delay or prevent delivery of defibrillation therapy to a victim suffering from sudden cardiac arrest.

Recalling firm
Zoll Medical Corporation
Classification
Class II
Status
Ongoing
Initiated
2024-05-31
Firm location
Deerfield, WI, United States

Product description

ZOLL Powerheart G5 AED, Semi-Automatic, G5Sxxx Family -Automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). Model/Catalog Number: G5S-00A G5S-00A-TSO G5S-00C G5S-01A G5S-02A G5S-02C G5S-02-L G5S-04A G5S-05A G5S-05C G5S-06A G5S-06C G5S-08A G5S-08C G5S-10A G5S-10C G5S-11A G5S-11C G5S-12C G5S-13A G5S-14A G5S-15A G5S-17A G5S-17C G5S-19A G5S-23C G5S-29A G5S-29C G5S-31A G5S-31A-SJA G5S-31C-SJA G5S-36A G5S-37A G5S-41A G5S-41C G5S-80A G5S-80A-TSO G5S-80C G5S-80-L G5S-82A G5S-82C G5S-83C G5S-90C

Reason for recall

G5 Semi-Automatic AED is shipped, with a protective film over its front panel, to protect the screen and shock button from cosmetic damage during shipping. It has come to our attention that customers may not be removing the protective film during deployment of the product. If the protective film is left adhered to the front bezel it may prevent the user from actuating the shock button and may lead to a delay or prevent delivery of defibrillation therapy to a victim suffering from sudden cardiac arrest

Data sourced from openFDA. This site is unofficial and independent of the FDA.