Recall Z-2321-2015— Siemens Medical Solutions USA, Inc.
Class II · Terminated
Class II FDA medical device recall by Siemens Medical Solutions USA, Inc., initiated 2015-06-25, terminated 2017-10-19. It names one FDA 510(k) clearance: K132804. Product: ACUSON S2000 and ACUSON S3000 ultrasound systems with the Virtual Touch IQ option; Model numbers: 10041461, 10440017 -S2000 system 10441730 -S3000 system VTIQ Option- 10439521 and 10439522 Radiology: The ultrasound imaging systems are intended for the following applications: Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.. Reason: Potential measurement error on ACUSON S Family ultrasound system. When repositioning the Virtual Touch IQ region of interest from the original (default) location, the lateral position of the measurement tool may not align with the lateral position of the shear velocity data..
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2015-06-25
- Terminated
- 2017-10-19
- Firm location
- Mountain View, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
ACUSON S2000 and ACUSON S3000 ultrasound systems with the Virtual Touch IQ option; Model numbers: 10041461, 10440017 -S2000 system 10441730 -S3000 system VTIQ Option- 10439521 and 10439522 Radiology: The ultrasound imaging systems are intended for the following applications: Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.
Reason for recall
Potential measurement error on ACUSON S Family ultrasound system. When repositioning the Virtual Touch IQ region of interest from the original (default) location, the lateral position of the measurement tool may not align with the lateral position of the shear velocity data.