Recall Z-2321-2012— AmeriWater Inc

Class II · Terminated

Class II FDA medical device recall by AmeriWater Inc, initiated 2012-08-09, terminated 2014-04-17. It names one FDA 510(k) clearance: K111740. Product: AmeriWater MRO Dialysis RO+ System, AmeriWater 1303 Stanley Avenue, Dayton, OH 45404 (800) 535-5585 www.AMERIWATER.com. The AmeriWater MRO Portable Reverse Osmosis Systems are water treatment systems intended for use in hemodialysis applications. They are designed to pre-treat and purify potable water for use in making dialysate for hemodialysis and to meet current AAMI and Federal (U.S.) standards. The AmeriWater Portable MROS model is intended for use in a hospital, clinic, dialysis center, or for home care for single patient use. The AmeriWater Portable MRO1 model is for treatment of up to two patients in a hospital, clinic, or dialysis centers.. Reason: The firm was notified that a malfunction of the device was caused by an incorrect fuse 32 volt 15 amp fuses instead of 250 volt 15 amp fuses, that was installed in the water system. The incorrect fuse may result in excessive heat in the fuse holder eventually resulting in failure of the device to operate..

Recalling firm
AmeriWater Inc
Classification
Class II
Status
Terminated
Initiated
2012-08-09
Terminated
2014-04-17
Firm location
Dayton, OH, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

AmeriWater MRO Dialysis RO+ System, AmeriWater 1303 Stanley Avenue, Dayton, OH 45404 (800) 535-5585 www.AMERIWATER.com. The AmeriWater MRO Portable Reverse Osmosis Systems are water treatment systems intended for use in hemodialysis applications. They are designed to pre-treat and purify potable water for use in making dialysate for hemodialysis and to meet current AAMI and Federal (U.S.) standards. The AmeriWater Portable MROS model is intended for use in a hospital, clinic, dialysis center, or for home care for single patient use. The AmeriWater Portable MRO1 model is for treatment of up to two patients in a hospital, clinic, or dialysis centers.

Reason for recall

The firm was notified that a malfunction of the device was caused by an incorrect fuse 32 volt 15 amp fuses instead of 250 volt 15 amp fuses, that was installed in the water system. The incorrect fuse may result in excessive heat in the fuse holder eventually resulting in failure of the device to operate.