Recall Z-2310-2016— Beckman Coulter Inc.

Class II · Terminated

Class II FDA medical device recall by Beckman Coulter Inc., initiated 2015-07-30, terminated 2017-02-15. It names one FDA 510(k) clearance: K112412. Product: AU5800 Chemistry Analyzers, Reference No. A98352 The AU series are automated, clinical chemistry analyzers that measure analytes in patient samples (serum, plasma, urine, CSF) in combination with appropriate reagents, calibrators, quality, control (QC) materials, and other accessories.. Reason: Beckman coulter has received complaints regarding leaking and cracked C cups. When the cup cracks, coolant can leak onto the connectors of the Printed Circuit Boards (PCB) in the analyzer, resulting in a short-circuit causing the instrument to go into a stop mode..

Recalling firm
Beckman Coulter Inc.
Classification
Class II
Status
Terminated
Initiated
2015-07-30
Terminated
2017-02-15
Firm location
Brea, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

AU5800 Chemistry Analyzers, Reference No. A98352 The AU series are automated, clinical chemistry analyzers that measure analytes in patient samples (serum, plasma, urine, CSF) in combination with appropriate reagents, calibrators, quality, control (QC) materials, and other accessories.

Reason for recall

Beckman coulter has received complaints regarding leaking and cracked C cups. When the cup cracks, coolant can leak onto the connectors of the Printed Circuit Boards (PCB) in the analyzer, resulting in a short-circuit causing the instrument to go into a stop mode.

Data sourced from openFDA. This site is unofficial and independent of the FDA.