Recall Z-2308-2019— Becton Dickinson & Company

Class II · Terminated

Class II FDA medical device recall by Becton Dickinson & Company, initiated 2019-06-12, terminated 2020-06-29. It names one FDA 510(k) clearance: K945952. Product: BD Vacutainer LH (Lithium Heparin) 34 I.U. Plus Blood Collection Tubes, Catalog Number 368494. Reason: False elevation of carboxyhemoglobin (COHb) results have been reported when using venous blood gas samples collected with tubes analyzed with the IL GEM 4000 instrument..

Recalling firm
Becton Dickinson & Company
Classification
Class II
Status
Terminated
Initiated
2019-06-12
Terminated
2020-06-29
Firm location
Franklin Lakes, NJ, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

BD Vacutainer LH (Lithium Heparin) 34 I.U. Plus Blood Collection Tubes, Catalog Number 368494

Reason for recall

False elevation of carboxyhemoglobin (COHb) results have been reported when using venous blood gas samples collected with tubes analyzed with the IL GEM 4000 instrument.

Data sourced from openFDA. This site is unofficial and independent of the FDA.