Recall Z-2307-2021— CooperSurgical, Inc.

Class II · Terminated

Class II FDA medical device recall by CooperSurgical, Inc., initiated 2021-06-30, terminated 2022-11-21. FDA's recall record names no specific 510(k) clearance for this event. Product: Oxygen Sensor Model K54019-2 used in the: G185 & G210 Incubators and G603 Work Chambers Oxygen Sensors supplied for use in the following incubator and work chamber products; G185 Standard 110V, G185 SensorTech 110V, G185 Standard 230V, G185 SensorTech 230V, G210 InviCell Standard, G210 InviCell Standard US, G210 Invicell Plus Optional pH and External Monitoring, G603 Work Chamber 115V, & G603 Work Chamber 230V. [CooperSurgical part numbers K22400-110, K22200-110, K22400-230, K22200-230, K59500, K59700, K60000, 3.641.831.153, & 3.641.832.153].. Reason: Units are displaying output readings that differ from the measurements taken using a Gas Analyzer. When the display is reading a target setting of 5%, an external meter is reading between 3% and 12% and the Incubator does not alarm. May result in loss of embryo during storage..

Recalling firm
CooperSurgical, Inc.
Classification
Class II
Status
Terminated
Initiated
2021-06-30
Terminated
2022-11-21
Firm location
Trumbull, CT, United States

Product description

Oxygen Sensor Model K54019-2 used in the: G185 & G210 Incubators and G603 Work Chambers Oxygen Sensors supplied for use in the following incubator and work chamber products; G185 Standard 110V, G185 SensorTech 110V, G185 Standard 230V, G185 SensorTech 230V, G210 InviCell Standard, G210 InviCell Standard US, G210 Invicell Plus Optional pH and External Monitoring, G603 Work Chamber 115V, & G603 Work Chamber 230V. [CooperSurgical part numbers K22400-110, K22200-110, K22400-230, K22200-230, K59500, K59700, K60000, 3.641.831.153, & 3.641.832.153].

Reason for recall

Units are displaying output readings that differ from the measurements taken using a Gas Analyzer. When the display is reading a target setting of 5%, an external meter is reading between 3% and 12% and the Incubator does not alarm. May result in loss of embryo during storage.

Data sourced from openFDA. This site is unofficial and independent of the FDA.