Recall Z-2306-2021— Philips Ultrasound Inc
Class II · Completed
Class II FDA medical device recall by Philips Ultrasound Inc, initiated 2021-07-22. It names one FDA 510(k) clearance: K182857. Product: EPIQ Ultrasound Systems, Model: EPIQ Elite, EPIQ 5G, EPIQ 5C, EPIQ 5W, EPIQ 7G, EPIQ 7C, EPIQ 7W, EPIQ CVx, EPIQ CVxi. Reason: Ultrasound system control panel arm swivel lock could fail, which could cause the control panel and monitor to swivel freely while a user is positioning the monitor, or transporting/moving the system, which could lead to patient or user musculoskeletal injury..
- Recalling firm
- Philips Ultrasound Inc
- Classification
- Class II
- Status
- Completed
- Initiated
- 2021-07-22
- Firm location
- Bothell, WA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
EPIQ Ultrasound Systems, Model: EPIQ Elite, EPIQ 5G, EPIQ 5C, EPIQ 5W, EPIQ 7G, EPIQ 7C, EPIQ 7W, EPIQ CVx, EPIQ CVxi
Reason for recall
Ultrasound system control panel arm swivel lock could fail, which could cause the control panel and monitor to swivel freely while a user is positioning the monitor, or transporting/moving the system, which could lead to patient or user musculoskeletal injury.
Data sourced from openFDA. This site is unofficial and independent of the FDA.