Recall Z-2302-2018— Intersurgical Inc

Class II · Terminated

Class II FDA medical device recall by Intersurgical Inc, initiated 2018-04-25, terminated 2019-04-09. FDA's recall record names no specific 510(k) clearance for this event. Product: Flexible Smoothbore Breathing Circuit with Exhalation Port 22mm, Product No. 80062000. Reason: When assembling an HME or Filter on this breathing circuit the tubing may be pushed up and inadvertently cover or block the exhalation port. This could potentially pose a serious health risk to the patient..

Recalling firm
Intersurgical Inc
Classification
Class II
Status
Terminated
Initiated
2018-04-25
Terminated
2019-04-09
Firm location
E. Syracuse, NY, United States

Product description

Flexible Smoothbore Breathing Circuit with Exhalation Port 22mm, Product No. 80062000

Reason for recall

When assembling an HME or Filter on this breathing circuit the tubing may be pushed up and inadvertently cover or block the exhalation port. This could potentially pose a serious health risk to the patient.